906-0731-01-05 Rev. C ZOLL Ventilator Operator’s Guide 4-23
Alarm Cancellation in Alarm Configuration Menu
There are clinical situations where an alarm occurs, and in the operators clinical judgment, this
alarm should be canceled for the remainder of the unit’s operating session. The following
constraints apply to alarm cancellation:
1. Only alarms that have occurred in the current operating session can be canceled.
2. Alarms which have not occurred since turn on are indicated with a “--”.
3. Canceled alarms are not be saved in the User Settings for the next session.
4. All canceled alarms reappear (if appropriate) when the unit is next turned on. (As an
example, the Self Check Fault, calibration due Alarm # 3120, reappears in the next
operating session.)
You may cancel the following alarms in the Alarm Configuration Menu:
1. Self Check Fault, calibration due (Alarm #3120)
2. RTC Battery Fault (Battery Low) (Alarm #3110)
3. Incomplete Exhalation (Alarm #3110)
4. PEEP Leak (Alarm #2090)
5. Fresh Gas Intake Fault (Alarm #3031)
Patient Inspiratory Demand Not Met (Alarm #3092)