SETTING UP THE ZOLL VENTILATOR
3-8 www.zoll.com 906-0731-01-05 Rev. C
7. Select Start Up Default Values
When you power on the ventilator, the Start Menu appears, through which you choose the
appropriate parameter default values for the patient. You can select the following patient
parameter defaults:
Adult
Pediatric
Mask CPAP -- Continuous Positive Airway Pressure (CPAP)
Custom -- Values saved in a previous session
Last Settings -- Values set for the patient last using the device before powering down
Note: The ventilator can may be configured to automatically select the Adult parameter
defaults at start up.
Adult Default Parameter Values
The Adult default parameter values are as follows:
Adult Parameter Default Values
Pediatric Default Parameter Values
The Pediatric default parameter values are as follows:
Pediatric Parameter Start-Up Defaults
Mode AC (V)
BPM 12
I:E 1:3
VT 450
PEEP 5
PIP Limit 35
FIO2 21
Mode SIMV (P)
BPM 20
Ti 0:6
PIP 20
PEEP 4
PIP Limit 30
FIO2 21