EasyManua.ls Logo

ZOLL AEV - Spo2 - Using the Pulse Oximeter

ZOLL AEV
156 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
906-0731-01-05 Rev. C ZOLL Ventilator Operator’s Guide 4-13
SpO2 -- Using the Pulse Oximeter
The primary use of the device is as a ventilator -- the pulse oximeter operates only when the
device is providing ventilation.
The following conditions can affect the pulse oximeter reading:
The sensor is too tight.
There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp,
or sunlight.
A blood pressure cuff is inflated on the same extremity as the one with a SpO2 sensor
attached.
The patient has hypotension, severe vascoconstriction, severe anemia, or hypothermia.
There is an arterial occlusion proximal to the sensor.
The patient is in cardiac arrest or is in shock.
The SpO
2
display is active only when the pulse oximeter is connected. The pulse oximeter is in
standby (and displays stby in the parameter window) when
No SpO
2
sensor is connected
The sensor is off the patient during start up
You place the pulse oximeter in standby
Note: You can place the pulse oximeter in standby only when the probe is disconnected from
the patient. A valid signal automatically brings the pulse oximeter out of standby.

Table of Contents

Related product manuals