906-0731-01-05 Rev. C ZOLL Ventilator Operator’s Guide 4-7
BPM Parameter Settings
The following table gives the options and ranges for the BPM parameters:
Parameter Window Options /
Range
Availability/Notes Models
Primary Value BPM
Breaths per
minute
0 to 80 Volume Target: Control Setting
Pressure Target: Measured
All
Secondary
parameter Val-
ues
Ti (sec)
|or
I:E
Ti 0.1 to 3.0
or
1:1 to 1:99
See BPM context menu: Con-
trol Parameter
All
Ti 0.1 to 5.0
or
I:E 4:1 to 1:99
Inverse I:E EMV+,
Eagle II
Alarm Limits High breath rate 20 to 99, off
Low breath rate 2 to 40
Measured Value Minute Volume
Vmin (ml)
0 to 99.9
BPM Context Menu
Control
Parameter
Default I:E 1:1 to
1:99, 0.3 to 3
The control value is shown in
the Parameter window, the
dependent value is shown in the
Auxiliary Box.
All
Ti 0.1 to 5.0
or
I:E $:1 to !:99
Inverse I:E EMV+,
Eagle II
Rise Time - Default # 1 to 10 Auxiliary Box All
Cycle Off %
(% Cycle )
Default 25% 10 to 70% Auxiliary Box All
Spont Ti Limit Default
Adult = 3.00
Infant = 2.00
Mask CPAP =
3.00
0.30 to 4.00