8
ENGLISH
System S2P (pod) only
WARNING
Always place the scanner in its intended mount on the
pod when not in use.
Do not place the pod on a slope greater than 5 degrees.
5.3 Explosion Hazards
Environment
WARNING
The TRIOS IOS system is not designed to be used in
environments that are potentially explosive such as
in close proximity to flammable liquids or gases or
in oxygen-enriched atmospheres.
Distance to the patient
WARNING
There is a potential explosion hazard if the TRIOS IOS system
is operated in the presence of flammable anesthetic.
5.4 Electrical Safety
For S2P (pod) only
PC connection options and requirements:
WARNING
The PC is standard IT equipment certified to IEC 60950-1
and is located at least 1.5 m away from the patient. If you
choose this setup, you must not scan the patient and touch
the PC at the same time! You must not touch any peripherals
connected with a cable to the PC while scanning a patient.
See the PC’s instructions for use for further details.
Ensure a grounded/earth connection
WARNING
To avoid the risk of electric shock, the TRIOS MOVE/
MOVE+ cart must only be connected to a supply mains with
grounding/protective earth. The TRIOS MOVE/MOVE+ cart
is equipped with a three pin (US: prongs) grounding/earth
type plug. If you cannot insert the plug into your outlet,
contact your electrician to replace the plug or outlet to
ensure a grounded power connection. Do not try to defeat
or ignore the purpose of the grounding-type plug.
Electrical shock
WARNING
There is a risk of electrical shock if you attempt to access
the inside of any part of the system. Only authorized and
qualified service personnel may access the inside of any
part of the system.
Do not use a power strip
WARNING
Do not connect the MOVE/MOVE+ cart to an ordinary power
strip or extension cord, because these may not be secure in
relation to grounding. If the ground connection fails, these
can cause the following hazards and potential injury:
The total leakage current for all the connected equipment
can exceed the limits specified in EN (IEC) 60601-1.
The impedance of the ground connection could exceed
the limits specified in EN (IEC) 60601-1.
Stress on cables
WARNING
All externally connected cables such as power cable,
network cable, must never be subjected to pulling stress.
Spilled liquids
WARNING
Do not bring liquids, such as beverages, near the TRIOS IOS
system. Do not spill liquids on the TRIOS IOS system.
Condensation
WARNING
Variations in temperature or humidity may cause water
to condense inside the system which can cause damage.
Always let the system reach room temperature before you
plug it into the power source.
If the system has been subject to major temperature or
humidity changes, wait at least 2 hours before plugging
it into the power source.
If there is visible evidence of condensation, wait at least
8 hours before plugging it into the power source.
Disconnecting from mains
WARNING
There is no power ON/OFF switch on the system;
therefore, the only reliable means to disconnect the system
from mains is to unplug the power cord. Do not position
the system so that it is difficult to unplug the power cord.
Please allow enough time for the system to shut down
properly before disconnecting the power cord.
USB service connection options and requirements
WARNING
Only connect bus-powered USB devices without
external power.