Safety
IM/265Gx/Ax-EN-07 265Gx, 265Ax 11
1.8.1 Compliance with Pressure Equipment Directive (97/23/EC)
Devices with PS > 200 bar (20 MPa)
Devices with a permissible pressure of PS > 200 bar (20 MPa) have been tested for conformity
by the Technical Supervisory Association TÜV NORD (0045) in accordance with module H and
can be used for fluids of group 1 (PED: 1G).
The name plate bears the following identification codes:
2
D-32425 Minden
0045
/M 100%/+ 0%/-
SIL
M00077
Type
Opt.
S/N
PED: 1G PS: 600 bar
265GS VEBNB1
E1L1B2CBCLH3
265GS6502016232
HART
SW xx
Bj./Yr. 20xx
IP 67
10,5 ... 30 V DC
4 ... 20 mA
Mat. 1.4404(8)/ AU/Hast.C / L
-1 ...+600 bar min.Span 6.0 bar
eing./adj.
+0.00 ... +50.000 bar
+0.00 ... +50.00 bar
MADE IN GERMANY
TAG: -/-
1
2
0044
II 1/2G Ex ia IIC T6 or T4
II 1D Ex iaD 20 T50°C or T95°C
Ta = -40°C to +40°C or +85°C
ZELM 01 ATEX 0064
Electr. Data see EC-Type-Examination Certificate
Fig. 3
1 PED: 1G (at PS > 20 MPa)
SEP (at PS 20 MPa)
2 CE 0045: Identification number of the
notified body with reference to the
Pressure Equipment Directive (for
PS > 20 MPa only)
Devices with PS 200 bar (20 MPa)
Devices with a permissible pressure of PS 200 bar (20 MPa) conform to Sec. 3 Para. (3) and
have not been tested for conformity. The devices have been constructed and manufactured
according to sound engineering practice (SEP).
The CE marking on the device does not refer to the Pressure Equipment Directive.
The name plate thus bears the identification code PED: SEP.