EasyManua.ls Logo

ABB Aztec 600

ABB Aztec 600
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers 3Installation
14 OI/AXM630–EN Rev. I
3.5.1 Connections Overview
Fig. 3.5 Connections Overview
+
+
+
+
+
+
AC Power Supply Earth (Ground) Stud
(on AC models only)
– see Warnings in Section 3.5, page 13
O/P 6
O/ P 5
O/P 4
O/P 3
O/P 2
O/P 1
Sockets for
Optional Digital
Communications
Ethernet RJ45
Connector
Socket
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
Attention
Relay
Failure Relay
Alarm 2 Relay
Alarm 4 Relay
Alarm 6 Relay
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
Stop
Relay
Calibration in
Progress Relay
Alarm 1 Relay
Alarm 3 Relay
Alarm 5 Relay
Fuse 1 A Type T
**Fuse 12.5 A Type T
L
N
E
+
E
OR
Use fuse rating:
AC Supply: 1 A (max.), Type T, 250 V
*DC Supply: 12.5 A 125 V DC Type T
*Manufacturer: SCHURTER,
Model SPT 5 x 20 Series
100 V to 240 V AC ±10 %
(90 V min. to 264 V max.)
50 / 60 Hz
18 to 36 V DC
** see Section 13.5,
page 69, for DC fuse
replacement details
TB7
Current Outputs
TB8
Optional Digital
Communications
TB2
TB3
TB4
TB5
TB6
TB1

Table of Contents