Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers 6 Configuration
34 OI/AXM630–EN Rev. I
6.2 Measurement
There are two tabs in the Measurement screen:
6.2.1 Setup
6.2.2 Streams
Fields Description
Chemical Units For certain parameters there is a choice of units to display the results.
Measuring Units The results can be expressed in a variety of units, for example by weight (mg or µg) or by volume
(ppm or ppb).
Temperature Units The results can be expressed in either Celsius (°C) or Fahrenheit (°F).
Flowcell and Heated Block
Temperature
The flowcell and heated block are heated and can be controlled at temperatures between 25 and
50 °C (77 and 122 °F).
Default temperatures are:
Ammonia – 35 °C (95 °F)
Fluoride – 30 °C (86 °F)
Fields Description
Stream 1
The Stream tag text is displayed in the operator views. Up to 20 characters can be used.
The Stream tag is also displayed in the configuration files and audit log files.