Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers 6 Configuration
OI/AXM630–EN Rev. I 43
6.8 Commissioning
There is one tab in the Commissioning screen:
6.8.1 Setup
Note. Access to the Commissioning Level is protected by an Internal Security Switch. The switch is, by default, set to 'Disabled'
and should be set to 'Enabled' only if the measurement parameter is to be changed. For further information on the function of the
Internal Security Switch, refer to Section 6.1.4, page 30.
Fields Description
Instrument type
The parameter to be measured by the analyzer. The options for ISE analyzers are:
Ammonia ISE
Fluoride ISE
Streams Fitted Fixed at 1.
Serial Number The serial number of the analyzer.
Mode of Operation
Options are:
Normal Run Mode
Demonstration Mode
Test M o de
Note.
Test mode is for diagnostic use only. Do not leave the analyzer running in this mode.
Before selecting 'Test Mode', disable archiving to SD card to prevent stored data conflicts.
Clear NonVol
Select 'Yes' to clear the analyzer's non-volatile memory.
Note. Selecting 'Yes' resets 'Instrument type' (see above) to its default setting of 'Iron monitor'
therefore the correct measurement parameter must be reselected.
Motor Speed Default setting is 2.80 rpm
Economy Mode Reduces reagent consumption (thereby lowering running costs) at the expense of a slower
response time. The options are:
Off
2.0 rpm.