8.2.1 Program a MultiWave bolus
Press d to move to the MULTIWAVE BOLUS
menu.
Select with f. The BOLUS AMOUNT screen
appears.
Press a to increase or s to decrease the total
bolus amount.
With the first press of the a key, the bolus
duration of the last MultiWave bolus appears, or
when using your insulin pump for the first time,
a default value appears.
In addition, as you increase or decrease the total
bolus amount, the suggested immediate bolus