10.2.7 Error E7: ELECTRONIC ERROR
The safety system of your insulin pump continuously monitors the
functioning of your insulin pump. If an electronic error is discovered, the
insulin delivery is stopped and an error E7: ELECTRONIC ERROR occurs.
You cannot turn off and confirm an error E7: ELECTRONIC ERROR by
pressing f, remove the battery instead.
Remove or disconnect the infusion set from your
infusion site.
Insert a new battery into the pump.
Prime the infusion set.
Reconnect the infusion set (tubing) at your
infusion site or insert the new infusion set.
When finished, put your insulin pump into RUN.
If an error E7: ELECTRONIC ERROR occurs during programming
(i.e. basal rate, time and date) check your settings.
If the error E7: ELECTRONIC ERROR occurs again, contact your local
Customer Care Area for assistance. Contact your doctor or healthcare
team for an alternate therapy plan, if necessary.