10.2.6 Error E6: MECHANICAL ERROR
With every insulin delivery (at least every 3 minutes) and each time you
put your insulin pump into RUN, the safety system of your insulin pump
reviews all mechanical parameters. If a mechanical error is discovered
during this process, the insulin delivery is stopped and an error E6:
MECHANICAL ERROR occurs.
Press f twice to turn off and confirm the error.
Press f to get to the Quick Info Screen. Note
down the remaining cartridge content.
Remove or disconnect the infusion set from your
infusion site.
Remove the cartridge, adapter, infusion set
(tubing) and battery from your insulin pump.
Insert a new battery. Rewind the piston rod.
Start the CHANGE THE CARTRIDGE function.
When the cartridge volume is displayed, press or
scroll a or s to move the plunger rod
forward until it reaches the remaining cartridge
content, which was noted down.
Set up your insulin pump with the cartridge and
a new infusion set. Prime the infusion set.
Reconnect the infusion set (tubing) at your
infusion site.
When finished put your insulin pump into RUN.
Should the error E6: MECHANICAL ERROR occur again in spite of these
actions, contact your doctor or healthcare team for an alternate therapy
plan. Contact your local Customer Care Area for further assistance.
158
Alerts and errors

Table of Contents

Other manuals for Accu-Chek Spirit

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Spirit and is the answer not in the manual?

Accu-Chek Spirit Specifications

General IconGeneral
TypeInsulin Pump
ManufacturerRoche Diabetes Care
Insulin TypeRapid-acting insulin
Bolus TypesStandard bolus, Extended bolus, Multiwave bolus
Battery TypeAAA alkaline
Reservoir Capacity315 units
Bolus Dose Increments0.1 units
DisplayLCD
ConnectivityInfrared
Battery LifeApproximately 4-6 weeks

Related product manuals