3.4 KeyLock
Symbol Beep Display Comment
A sequence of beeps means that all If no KeyLock symbol
(you will hear 3 beeps 4 keys are locked appears in the RUN
in RUN and 1 beep in (except a for turning or STOP screen, the
STOP). on the backlight). KeyLock function is
turned off.
means that all
4 keys are unlocked.
The KeyLock function allows you to lock all four keys on your insulin
pump and serves as an additional safety measure against unintentional
activation of functions (e.g. while sleeping, during sports). To make use of
the KeyLock function, it must be turned on in the SETUP MENU
STANDARD.
If the KeyLock function is turned on, it needs to be unlocked or turned off
before any programming of your insulin pump can be performed. All
instructions in this user guide are given with the assumption that the
KeyLock is NOT turned on. To unlock the KeyLock, press d + s
simultaneously (you will hear three beeps in RUN, or one beep in STOP)
and hold for 3 seconds until you hear again three beeps (in RUN) or one
beep (in STOP).
Symbol Timeout g
For your safety and convenience during programming,
your insulin pump automatically returns to the RUN or
STOP screen if no key is pressed within 20 seconds h.
For your safety, the changes you made before your
insulin pump “times out” will not be saved.
Within this user guide the symbol g represents the
timeout feature.