10.2.8 Error E8: POWER INTERRUPT
A power interrupt may occur when you change the battery without
putting your insulin pump into STOP first, or when your insulin pump
was dropped.
Press f twice to turn off and confirm the error.
Your insulin pump goes into STOP.
Check the time and date and correct,
if necessary.
If a bolus and/or a temporary basal rate was interrupted by the error,
review the bolus and/or the temporary basal rate history for the delivered
amount and duration.
When finished, put your insulin pump into RUN,
if necessary.
If necessary, program a new bolus and/or
temporary basal rate.
If an error E8: POWER INTERRUPT occurs during programming, check
your settings.
160
Alerts and errors

Table of Contents