3.9 STOP-Warning
The STOP-Warning is turned on each time your pump is changed from
RUN into STOP or when a battery is inserted. This reminder is to inform
you that insulin delivery is currently interrupted. If the STOP-Warning is
turned on, it occurs every minute with a long beep and a vibration.
Turning off the STOP-Warning stops this warning function.
To turn off the STOP-Warning
Action Result
Press and hold You hear a melody. If you press a or s too
a or s The STOP-Warning briefly, the STOP-Warning
(for 3 seconds). is turned off. is not suppressed and occurs again.
The volume of the STOP-Warning is
independent of the programmed beep
volume. It always occurs at the
maximum level.
The STOP-Warning will be reactivated the next time your pump is
changed from RUN into STOP or with the next battery change.
The operation of your insulin pump is logical and simple. For your
convenience a “map” of these functions is found in the back of this user
guide.