Alerts and errors
10.2.10 Error E11: SET NOT PRIMED
If you changed the cartridge and infusion set but did not prime the
infusion set, an error E11: SET NOT PRIMED occurs when you try to put
your insulin pump into RUN.
Press f twice to turn off and confirm the error.
Remove or disconnect the infusion set from your
infusion site.
Prime the infusion set.
Reconnect the infusion set (tubing) at your
infusion site or insert the new infusion set.
When finished, put your insulin pump into RUN.
Warning
Never prime an infusion set that is connected to your
body. You risk uncontrolled insulin delivery into your
body. With disconnectable infusion sets, make sure you
have disconnected the tubing from your site prior to
priming. Always follow the instructions provided with
your infusion set.

Table of Contents

Other manuals for Accu-Chek Spirit

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Spirit and is the answer not in the manual?

Accu-Chek Spirit Specifications

General IconGeneral
TypeInsulin Pump
ManufacturerRoche Diabetes Care
Insulin TypeRapid-acting insulin
Bolus TypesStandard bolus, Extended bolus, Multiwave bolus
Battery TypeAAA alkaline
Reservoir Capacity315 units
Bolus Dose Increments0.1 units
DisplayLCD
ConnectivityInfrared
Battery LifeApproximately 4-6 weeks

Related product manuals