determine the timing as well as the amount and type of bolus you
need to deliver. Make sure you know your personal carbohydrate
insulin ratio and your correction bolus ratio.
➨ When your insulin pump is in STOP, it does not deliver any
insulin. Put your insulin pump into RUN to continue the insulin
delivery.
➨ After inserting a new battery, always ensure that the time
and date of your insulin pump are set correctly. Incorrect
programming of the time and date may cause incorrect insulin
delivery.
➨ If your insulin pump “times out” before you finish
programming a bolus, no bolus is delivered. Check the bolus
history screens on your insulin pump for the bolus amount
delivered with time and date and program a new bolus if necessary.
➨ If you do not set the correct battery type on your insulin pump,
the alert A2: BATTERY LOW may not occur in time to provide you
sufficient warning to replace your battery.
➨ If you flip the display orientation of your insulin pump by
180°, the a and s keys will also reverse their functions. In
comparison to the display orientation, the upper key will be a
and the lower key will be s. This change applies for all a and
s functions including turning on the backlight. The d and f
keys will not change in their function, but will remain the same,
regardless of your screen orientation.
➨ If data transfer between your PC and your insulin pump is
disrupted in any way, the configuration may be incomplete and an
error E12: DATA INTERRUPTED occurs. The data transfer must be
completed successfully before you can put your insulin pump into
RUN.