Make sure your insulin pump is in STOP.
Press d to move to the PRIME THE INFUSION
SET menu.
Select with f.
START PRIMING and the priming amount of
25 units h of insulin is shown on the display.
Press and hold f until you hear a melody and
the piston rod moves forward. The priming
process starts and a count-up is shown on the
display. Your insulin pump automatically stops
priming after 25 units h of insulin.
When the priming process is finished, your
insulin pump returns to STOP.
The infusion set is properly primed when no air
bubbles are visible in the tubing anymore and a
bubble-free flow of insulin emerges from the tip
of the needle.
start
priming
start-up
procedure
stop
priming
STOP
screen
prime the
infusion set