Note When an Extended bolus, a MultiWave bolus and/or a tempo-
rary basal rate (TBR) is/are active and you put your insulin
pump into STOP, an alert A6: TBR CANCELLED and/or an
alert A8: BOLUS CANCELLED occur(s).
Press f twice for each appearing alert to confirm and
turn off.
See the sections 10.1.6 “Alert A6: TBR CANCELLED”
and 10.1.8 “Alert A8: BOLUS CANCELLED” for further
information.
Turn off the STOP-Warning
Press and hold s or a until you hear a melody. The STOP-Warning is
turned off.
Note The volume of the STOP-Warning is independent of the
programmed beep volume. It always occurs at the maximum
level.
The STOP-Warning will be reactivated the next time your pump is
changed from RUN to STOP or with the next battery change.
STOP-Warning
The STOP-Warning is turned on each time your insulin
pump is put from RUN into STOP or when a battery is
inserted. To inform you that the insulin delivery is inter-
rupted, your insulin pump reminds you every minute
with a long beep and a vibration. Turning off the
STOP-Warning suppresses this warning function.