121
When J is in low state as shown in Fig 5-8, OUT is in low state. When J is in high
state, OUT is in high state. OUT can therefore output pulse signals under the
control of J.
The max output voltage and current of digital output circuit are 250V and
100mA respectively.
Another drawing of the alarming output with buzzer is shown in Fig 5-9.
Fig 5-9 schematic diagram of digital output circuit 2
The wire of the digital output circuit should be chosen between AWG22~16 or
0.5~1.3 mm
2
.
Wiring of Analog Output Circuit:
There are 2 analog output circuits in AXM-IO2 modules. The terminals of the
analog output circuits are AO1+, AO1- and AO2+, AO2-. The analog output
circuit can convert any one of 30 electrical quantities, which is selected by
user, to analog voltage or current. The analog output circuit supplies 4 output
modes, including 0 to 20mA mode, 4 to 20mA mode, 0 to 5V mode and 1 to 5V
mode.