EasyManuals Logo

AccuEnergy Acuvim II series User Manual

AccuEnergy Acuvim II series
350 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #257 background imageLoading...
Page #257 background image
249
Address Parameter
Format
code
Range
Data
type
Property
104FH
First group: comparison
mode
F1
1: greater than; 2:
equal to; 3: less
than
word R/W
1050H First group: setpoint value F10~F18
Related with
parameters
word R/W
1051H First group: delay F1 0~3000(*10ms) word R/W
1052H First group: output to relay F1
0:none,
1~8: related relay
word R/W
1053H~
109DH
2
nd
to 16
th
group
Same as the
rst group
word R/W
Alarming parameter code table
Table 6-20
Setting
value
Alarming object
Setting
value
Alarming object
Setting
value
Alarming object
0 Frequency 1 Va 2 Vb
3 Vc 4
Average phase
voltage
5 Uab
6 Ubc 7 Uca 8 Average line voltage
9
current of phase
A
10 current of phase B 11 current of phase C
12 Average current 13 Neutral current 14 Power of phase A
15 Power of phase B 16 Power of phase C 17 Power of all
18
Reactive power of
phase A
19
Reactive power of
phase B
20
Reactive power of
phase C
21
Reactive power of
all
22
Apparent power
of phase A
23
Apparent power of
phase B
24
Apparent power
of phase C
25
Apparent power
of all
26 PF of A
27 PF of B 28 PF of C 29 PF
30
Voltage
unbalance factor
U_unbl
31
Current unbalance
factor I_unbl
32
Load characteristic(R/
L/C)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AccuEnergy Acuvim II series and is the answer not in the manual?

AccuEnergy Acuvim II series Specifications

General IconGeneral
BrandAccuEnergy
ModelAcuvim II series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals