EasyManua.ls Logo

Acteon PSPIX - Page 42

Acteon PSPIX
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ENGLISH
PSPIX² • User manual
42
Disinfecting the PSPIX
system
Remove the removable parts
from the PSPIX system (IP
insertion part and receptacle)
and disinfect them separately.
-Disinfect the PSPIX body with
a cloth dipped in a disinfectant
solution.
All surfaces must be dry
before using the PSPIX
system.
-Never use disinfectants
that contain abrasive or
corrosive substances, or
that contain solvents.
-Never spray the
disinfectant directly
onto the PSPIX system.
-Do not introduce
disinfectant in the PSPIX
system’s insertion slot.
Never use a wet cloth on
the black part under the
IP’s insertion slot.
Disinfection of
removable parts
(IP insertion part and
receptacle)
Remove the removable parts
from the PSPIX system (IP
insertion part and receptacle)
To disinfect the removable
parts, use
-a cloth dipped in a disinfectant
solution.
-or washer disinfector. The
following logo
on the
back of the removable parts
indicates that they are thIPo-
disinfectable.
Program your washer disinfector
so that the thermal disinfection
takes place at 93 °C for ve
minutes.
-Ensure that the parts are
completely dry and are
at ambient temperature
before putting them back
into the PSPIX system and
using them.
-Never use disinfectants
that contain abrasive or
corrosive substances, or
that contain solvents.
-When using a washer
disinfector, do not
exceed a period of ve
minutes when thermal
disinfection takes place
at 93 °C.
https://stomshop.pro