USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
ENGLISH
TMJ nose support
This support must be used for TMJ examinations since it is specically designed to stabilize the patient’s head in both open
mouth and closed mouth conditions.
Calibration tray and geometric calibration phantom
The Calibration tray is used to position the geometric calibration phantom during the calibration procedure of the CBCT
apparatus.
The calibration procedure consists of a scan of the calibration phantom and the estimation of a set of parameters that des-
cribe the conguration of the system.
Quality phantom support for Germany
This accessory is intended to position the QUART DVT test phantoms for QA/QC of CBCT functionality for X-MIND trium
installed in Germany.
Self standing base
This accessory allows to install X-Mind trium where the wall cannot adequately support trium load.