USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
ENGLISH
- medium
- large
· WOMAN
- small
- medium
- large
· CHILD
- small
- medium
- large
For the panoramic group and the cephalometric group, the operator can still modify the loading factors by means of dedicated
sliders.
!
WARNING :
Only physicians or dentists properly trained can modify the loading factors from the predened values for the panoramic and
cephalometric groups taking into account the effects on the image quality, increase/decrease of radiation dose and the needs of the
clinical task.
On the contrary, for the CBCT group, the operator cannot modify the anodic current and tube voltage, but can just select the
quality of the 3D reconstruction between three protocols:
· Standard quality
· Medium quality
· High quality
The X-ray exposure time of the scan increases from the standard quality to the high quality involving both the improvement
of the image quality and the increase of the radiation dose to the patient.
!
WARNING :
Only physicians or dentists properly trained can select the quality protocol for the CBCT groups taking into account the effects on the
image quality, increase / decrease of radiation dose and the needs of the clinical task.
2.6 CONTROL PANELS
Equipment control panels are used during patient positioning phase.
Available keys allow to move the equipment to Initial position before patient access, turn on positioning LASERs, adjust the
U-arm position to align the patient to X-ray beam and set the “ready to X-ray” status.
Some keys are duplicated on the CEPH control panel, for use during patient positioning for CEPH exams.
Equipment control panel CEPH control panel