99
44
USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
ENGLISH
MAINTENANCE, CLEANING AND
DISPOSAL
9.1 MAINTENANCE
Periodic verication of the emergency switches.
To ensure the safety of patients and operators, verify the proper operation of the remote emergency switch and local
emergency switch on a monthly basis.
To verify the proper operation of remote and local emergency switches, proceed as follows:
• Turn the device on and make sure that it is operating correctly.
• Click on the U-arm rotation button on the control panel to start a reset to 0 position of the U-arm.
• Whilst the U-Arm is in motion, press the local emergency switch; the U-Arm should immediately stop as a proof of the proper
operation of the emergency switch.
• Reset the emergency status by turning the local emergency switch clockwise and bringing it back to its initial position.
• Repeat the same procedure for the remote emergency switch.
CAUTION :
If an emergency switch does not work properly, contact the technical support service.
Calibration and quality control
Calibration and quality tests verify the operation and performance of the device; these tests must be carried out together
with the inspection of the safety devices at least once every 6 months, unless otherwise specied in the frequency table of the
installation and maintenance manual.
For other details of maintenance operations, refer to the installation and maintenance manuals.
The manufacturer shall not be held liable for damage or injuries caused by failure to carry out inspections and tests and by
incomplete maintenance.
Repairs and replacements of any component must be carried out solely by authorized and highly qualied personnel and only
using genuine spare parts supplied by de Götzen® S.r.l. - ACTEON Group.
9.2 CLEANING
Clean the external surface using a damp cloth with non-corrosive and non oil-based detergent; disinfect the external surface
using a non-aggressive medical detergent. Do not spray any detergent or disinfectant directly on the device.
9.3 DISPOSAL
The WEEE symbol indicates that, at the end of its lifespan, the product must be disposed of separately from other waste, in
compliance with Directive WEEE 2012/19 EU.
Refer to the implementation standards in your country. EU Council Directive WEEE 2012/19 EU denes a common approach
intended to avoid, prevent or reduce harmful effects due to the exposure to environmental noise and to the disposal of
electric and electronic equipment. This product is marked with the symbol shown above. This product must not be disposed
of together with domestic waste. It must be taken to a special waste collection centre to be recovered and recycled. The
crossed-out wheelie bin identies a product placed on the market after the 13th of August 2005 (see EN 50419:2006). This
product is subjected to Council Directive WEEE 2012/19 EU and national implementation standards. Refer to your supplier for
the disposal of this product.
Proper disposal of this product will help protect the environment.
For further details on the disposal of this product, please contact local authorities, the provider of the domestic waste disposal
service or the dealer where you have purchased it.