USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
Macbook PRO product line or
iMac product line with the following or equivalent characteristics:
RAM: minimum 4 GB, for big FOV DICOM stacks it is suggested to install 8 GB.
Hard disk: minimum 300 GB (for image storage, the software alone occupies 200 MB, the full implant library is 7 Gb size)
Graphic card for optimal performance: Nvidia Geforce product range or Nvidia Quadro with 1 Gb dedicated RAM.
Screen resolution: minimum height 1024 and width 1600, it is suggested to use higher resolutions for optimal planning (full
HD: 1920 x 1080)
O/S: OS X Sierra (10.12).
Network Cards : A LAN network must be used for the connection between client and server.
Avoid to use a wireless network card on the client : fast connection network (LAN type)
are needed to share the patient exams with the clients PCs. Due to this, a WiFi network
card could be too slow to transfer patient exams with clients PCs.
OTHER SOFTWARE :
• If you install an antivirus or rewall or internet security in the workstation, take care of these specic :
- All the AIS processes have to be excluded in AV the exclusion list
- All the ports used by AIS (from 51000 to 51020, and 6543) have to be opened in the rewall
- If you face other problems or you are not able to congure AV or rewall, please disable them.
PAY ATTENTION TO DON’T HAVE AIS WORKSTATION SOFTWARE VERSION ALREADY
INSTALLED IN THE MACHINE. OTHERWISE YOU WILL HAVE PROBLEM IN THE INSTAL-
LATION AND IN THE CONFIGURATION AIS CLIENT VERSION !
11.13 INTENDED ENVIRONMENT
Clinical environment conditions (use conditions):
• Temperature: 10 to 30°C
• Relative humidity: 25 to 75 %
• Atmospheric pressure: 850 to 1060 hPa
Transportation environment conditions:
• Temperature: 0 to 50°C
• Relative humidity: See clinical environment conditions
• Atmospheric pressure: 500 to 1060 hPa
Warehousing environment conditions:
• See Transportation environment conditions
11.14 DIMENSIONS OF THE UNIT
PAN/CBCT