EasyManua.ls Logo

ADCMT 7451A - Page 144

ADCMT 7451A
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-44
5. Status command
6. Common command to IEEE-488.2 (Can be used in USB.)
Function description SCPI Command Parameter Query reply
Operation Event Enable
Register setting
:STATus:OPERation:ENABle [0|0|65535] <int_5>
Operation Event Register reading
:STATus:OPERation:EVENt? <int_5>
Questionable Event Enable
Register setting
:STATus:QUEStionable:ENABle [0|0|65535] <int_5>
Questionable Event Register
reading
:STATus:QUEStionable:EVENt? <int_5>
Measurement Event Enable
Register setting
:STATus:MEASurement:ENABle n [0|0|65535] <int_5>
Measurement Event Register
reading
:STATus:MEASurement:EVENt? <int_5>
Function description SCPI Command Parameter Query reply
Each status byte clear *CLS
Standard Event Status Enable Reg-
ister setting
*ESE {0|0|127] <int_3>
Standard Event Status Register
reading
*ESR? <int_3>
Instrument information reading *IDN? For more information
on a query response,
refer to page 6-32.
Setting the operation complete bit
(bit0) of the Standard Event Status
Register after all operations are
complete
*OPC
Processing state of the operarion is
returned
*OPC? {0|1}
(0: completed,
1: uncompleted)
Instrument initialization *RST
Service Request Enable Register
setting
*SRE [64|64|127]
(bit6 is always 1)
<int_3>
Status Byte Register reading *STB? <int_3>
Trigger command *TRG
Self-test performing and the result
reading
*TST? {0|1}
(0: Normal, 1: Abnor-
mal)
All operation completion waiting *WAI (GPIB only)

Table of Contents

Related product manuals