4451C EN 2013-04-30Introducing the CR 85-X
CR 85-X DIGITIZER
When installing the CR 85-X, care must be taken to ensure that there is
either a mains plug or an all-cable disconnecting device in the internal
installation fitted near the CR 85-X and that it is easily accessible.
If you notice conspicuous noise or smoke, disconnect the CR 85-X
immediately.
Check that the mains voltage is within the specified range of the self
adapting power supply of the machine.
The classification of this product according to the medical electrical
equipment standard IEC 60601-1 requires installation outside the patient
vicinity. For definition of patient vicinity see dimensions below.
R = 1.5 m/4.9 feet (EN 60601-1) or 1.83 m/6 feet (UL 60601-1).
h = 2.5 m/8.2 feet (EN 60601-1) or 2.29 m/7.5 feet (UL 60601-1).
WARNING: To avoid risk of electric shock, this equipment must only
be connected to a supply mains with protective earth.
Patient vicinity
Patient vicinity