14
4451C EN 2013-04-30Introducing the CR 85-X
CR 85-X DIGITIZER
• When installing the CR 85-X, care must be taken to ensure that there is
either a mains plug or an all-cable disconnecting device in the internal
installation fitted near the CR 85-X and that it is easily accessible.
• If you notice conspicuous noise or smoke, disconnect the CR 85-X
immediately.
• Check that the mains voltage is within the specified range of the self
adapting power supply of the machine.
• The classification of this product according to the medical electrical
equipment standard IEC 60601-1 requires installation outside the patient
vicinity. For definition of patient vicinity see dimensions below.
R = 1.5 m/4.9 feet (EN 60601-1) or 1.83 m/6 feet (UL 60601-1).
h = 2.5 m/8.2 feet (EN 60601-1) or 2.29 m/7.5 feet (UL 60601-1).
WARNING: To avoid risk of electric shock, this equipment must only
be connected to a supply mains with protective earth.
Patient vicinity
Patient vicinity
R
h