4451C EN 2013-04-30 Advanced operation (‘Key-operator mode’)
CR 85-X DIGITIZER
8 Press the Confirm key to activate the new CPF.
Press the Confirm key to consult the image transmission queue and if necessary delete
the images in the queue. Refer to ‘Consulting the images in the queue’ on page
52.
When the queue is empty, restart the ‘Install - Configuration’ function, refer to step 2.
9 Wait until the CR 85-X displays:
This screen offers you the possibility to change the IP address of the CR 85-X.
10 You have the choice whether or not to change the IP address:
To keep the current IP address, press the Confirm key.
Note: If the CR 85-X displays the screen below, the image queue is not empty,
and therefore you cannot change the IP address of the CR 85-X.
New CPF detected . . .
Allocation / change of
name/ip-address?
: cancel
: ok
INSTALL
CONFIGURATION
SERVICE XXXXX
: quit
: ok
INSTALL
CONFIGURATION
SERVICE XXXXX
Image queue not empty!
Check the queue then
restart ‘INSTALL -
CONFIGURATION’ -
Press for queue
management . . .
: quit
: ok
: select
INSTALL
CONFIGURATION
SERVICE XXXXX
Confirm ID or select new:
SOLOxxxxxxxxxx1.............. 201
SOLOxxxxxxxxxx2 ............. 202
COMPxxxxxxxxxx3 ............ 203
SOLOxxxxxxxxxx4 ............. 204
COMPxxxxxxxxxx5............. 205
SOLOxxxxxxxxxx6.............. 206

Table of Contents

Other manuals for AGFA CR 85-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AGFA CR 85-X and is the answer not in the manual?

AGFA CR 85-X Specifications

General IconGeneral
BrandAGFA
ModelCR 85-X
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals