78
4451C EN 2013-04-30Advanced operation (‘Key-operator mode’)
CR 85-X DIGITIZER
• To change the IP address of the CR 85-X, select an IP address and the
corresponding name via the Up and Down keys and confirm.
The CR 85-X will display:
11 Wait until the CR 85-X displays:
You must make backup files with the new parameters.
12 Insert an empty formatted floppy into the disk drive and press the Confirm key.
The CR 85-X will start copying the machine specific configuration data while
displaying the Save configuration screen:
Initializing new
station name . . .
INSTALL
CONFIGURATION
SERVICE XXXXX
Configuration installed.
Parameters have changed.
You should refresh your
backup now.
Please insert the backup
floppy and press
: quit
: ok
INSTALL
CONFIGURATION
SERVICE XXXXX
copying . . .
D:\ACP_<serial#>.zip
to
A:<path><file name>
SAVE
CONFIGURATION
SERVICE XXXXX