Chapter 4 Calibration and Adjustment
-24 dB Range Flatness Adjustment
4
4
3 Using the numeric keypad or knob, adjust the displayed voltage at each setup
to match the measured voltage. Select ENTER VALUE.
4 After performing setup 45:
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform “-24 dB Range
Flatness Verification”, on page 101.
b. If you are making all the adjustments and then verifying the
instrument’s performance, continue with the next procedure in
this chapter.