EasyManua.ls Logo

Agilent Technologies 33500 Series - Page 119

Agilent Technologies 33500 Series
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
119
Chapter 4 Calibration and Adjustment
-8 dB Range Flatness Adjustment
4
4
3 Using the numeric keypad or knob, adjust the displayed voltage at each setup
to match the measured voltage. Select ENTER VALUE.
4 After performing setup 53:
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform “-8 dB Range
Flatness Verification”, on page 99.
b. If you are making all the adjustments and then verifying the
instrument’s performance, verify the output specifications of the
instrument using the “Performance Verification Tests”, on page
92.
This completes the adjustment procedures for the 35521A 1-Channel
Function/Arb Generator. Verification of the output specifications is
recommended.
If you are making adjustments to a 35522A 2-Channel Function/Arb
Generator, continue with the next procedure in this chapter.

Table of Contents

Other manuals for Agilent Technologies 33500 Series

Related product manuals