Amplitude Accuracy (continued)
Measurement Lower
Limit
Upper Limit
(dBfs)
Measured Value
(dBfs)
Pass/Fail
-35 dBVrms, Ch 2 -35.15 -34.85
-35 dBVrms, Ch 3 -35.15 -34.85
-35 dBVrms, Ch 4 -35.15 -34.85
-27 dBVrms, Ch 1 -27.15 -26.85
-27 dBVrms, Ch 2 -27.15 -26.85
-27 dBVrms, Ch 3 -27.15 -26.85
-27 dBVrms, Ch 4 -27.15 -26.85
-11 dBVrms, Ch 1 -11.15 -10.85
-11 dBVrms, Ch 2 -11.15 -10.85
-11 dBVrms, Ch 3 -11.15 -10.85
-11 dBVrms, Ch 4 -11.15 -10.85
1 dBVrms, Ch 1 0.85 1.15
1 dBVrms, Ch 2 0.85 1.15
1 dBVrms, Ch 3 0.85 1.15
1 dBVrms, Ch 4 0.85 1.15
9 dBVrms, Ch 1 8.85 9.15
9 dBVrms, Ch 2 8.85 9.15
9 dBVrms, Ch 3 8.85 9.15
9 dBVrms, Ch 4 8.85 9.15
19 dBVrms, Ch 1 18.85 19.15
19 dBVrms, Ch 2 18.85 19.15
19 dBVrms, Ch 3 18.85 19.15
19 dBVrms, Ch 4 18.85 19.15
27 dBVrms, Ch 1 26.85 27.15
27 dBVrms, Ch 2 26.85 27.15
27 dBVrms, Ch 3 26.85 27.15
27 dBVrms, Ch 4 26.85 27.15
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Verifying Specifications Agilent 35670A
Performance Test Record - Four Channel