ICP Supply
Measurement Lower
Limit
Upper Limit Measured Value Pass/Fail
Ch 1 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 2 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 31 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 4 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 1 Current 2.75 mA 5.75 mA mA
Ch 2 Current 2.75 mA 5.75 mA mA
Ch 3 Current 2.75 mA 5.75 mA mA
Ch 4 Current 2.75 mA 5.75 mA mA
Source Amplitude Accuracy
Measurement Lower
Limit
(%)
Upper Limit
(%)
Measured Value
(%)
Pass/Fail
1 kHz, 0.1 Vpk -4 4
1 kHz, 3.0 Vpk -4 4
1 kHz, 5.0 Vpk -4 4
Source Flatness
Measurement Lower
Limit
(dB)
Upper Limit
(dB)
Measured Value
(dB)
Pass/Fail
12.8 kHz -1 1
25.6 kHz -1 1
51.2 kHz -1 1
Source Distortion
Measurement Lower
Limit
Upper Limit
(dBc)
Measured Value
(dBc)
Pass/Fail
12.8 kHz -60
51.2 kHz -40
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Verifying Specifications Agilent 35670A
Operation Verification Test Record - Four Channel
14 of 14