Page #16 background image
Adjustment
Tools
No
hand
tools
are
required for
the adjustments
.
A
djustable Components
There
are
no
adjustable
components
, such
as variable
capacitors or
variable resistors
,
in
the
4263B
.
All
adjustments
are
performed by
updating the
adjustment data
stored in
the
internal
EEPROM.
F
actory-Selected Components
No
factory-selected
components
are used
in the
4263B
EEPROM
W
rite
Protect
Switch
There
are no
EEPROM
write
protect
switches
in
the
4263B
.
Related
A
djustments
There
are
two
adjustments
for
the
4263B
.
T
able
2-1
lists
the
adjustments
that
must
be
performed
if
an
assembly
has
been
replaced.
T
able
2-1.
Related
A
djustments
Assembly
Replaced
Required
A
djustment
A1
Main Board
Assy
All
adjustments
A2
CPU
Board
Assy
without
ROM
None
1
A5
A
C
Inlet
Board
Assy
None
LCD Assembly None
DC-DC Converter Unit None
Keyboard Assembly None
1
If the EEPROM, A2U1, is replaced, all adjustments are required.
2-2 Adjustments

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B LCR and is the answer not in the manual?

Agilent Technologies 4263B LCR Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B LCR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals