Agilent Technologies 4263B LCR - Impedance Measurement 1 m Adjustment; Equipment; Setup

Agilent Technologies 4263B LCR
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Impedance
Measurement
1m
A
djustment
This
adjustment
updates
the
data
to
adjust
the
measurement
impedance
to the
standard values
,
when
the
1m
test
leads
are
used.
Note
P
erform the
Impedance Measurement
0m
A
djustment
prior
to
performing
this
adjustment.
Equipment
Standard
Resistors
42030A
Open
T
ermination
42090A
Short
T
ermination
42091A
Test
Leads (
1m
)
16048A
Setup
This
setup
obtains
the
impedance
measurement
adjustment
data
for
1
m
test
leads
.
Figure 2-4. Impedance Measurement 1m Adjustment Setup
Perform the following steps as the program instructs.
1. Connect the 16048A directly to the 4263B UNKNOWN Connectors.
Adjustments 2-9