Transducer
Section
The
transducer section
measures
the
vector
voltage
across
the
DUT
,
and
the
current
owing
through
the DUT
by
converting
the
current
to
voltage
.
The
transducer
also
includes
a switch
used
by the
Option
001
(Transformer
P
arameter
Measurement),
and
the
contact
check
voltage
source
.
Figure
4-5
shows
the
transducer
block
diagram.
Figure
4-5.
Transducer
Block
Diagram
The
voltage
across
the
DUT
is
measured
at
the
H
POT
and
L
POT
connectors
using
a
dierential
amplier
.
The
current
owing
through
the
DUT
and
owing
into
the
L
CUR
connector
is
converted
to
voltage
by
an
I-V
converter
.
The
dierential
amplier
measures
the
voltage
between an
I-V
converter
output
and
the inverting
input of
the converter
.
The
inverting
input
of
the
converter
,
thus
L
CUR
,
is
at
virtual
ground, because
of the
very high
input impedance
and
very
high
open
loop
gain
of the
operational amplier
used in
the I-V
converter
.
Troubleshooting 4-7

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B LCR and is the answer not in the manual?

Agilent Technologies 4263B LCR Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B LCR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals