Theory of
Operation
This
section
describes
the
general overall
operation of
the 4263B
and the
operation of
each
assembly
.
Overall
Measurement
Theory
The
4263B measures
the
impedance
of
the
Device
Under
T
est
(DUT)
by
measuring
vector
voltage
to current
ratio
.
A
vector
is
a
value
which
consists
of
a
magnitude
and
a
phase
.
The
impedance
of
the
Device
Under
T
est
(DUT)
is
determined
by
the
vector
ratio
between
the
voltage
across
the
DUT
and
the
current
owing
thorough
it.
The
4263B
essentially
consists
of
a
signal
source
,
a
vector
voltmeter
,
and
a
vector
current
meter
.
Figure
4-1. V
oltage Current
Ratio Measurement
Principle
In
Figure 4-1
,a
DUT connected
in
series
with
a
test
voltage
source
(including
an
output
resistance),
voltmeter
V
,
and
ammeter
,
A.
If
the
ammeter
measures
i
x
ampere
when
the
source
voltage
is
e
x
,
the DUT's
impedance
Zx
is
expressed
by:
Zx
=
e
x
i
x
Note
In this chapter
, we will
focus only on
impedance
measurement. The other
parameters, such as L, C, and R, are derived mathematically from the measured
impedance value
.
Troubleshooting 4-3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B LCR and is the answer not in the manual?

Agilent Technologies 4263B LCR Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B LCR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals