Agilent Technologies 4263B LCR - Check 11, DC-DC Converter Open Output Voltages

Agilent Technologies 4263B LCR
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Check
11, DC-DC
Converter Open
Output
V
oltages
1.
Turn o
the
4263B
.
2.
Disconnect the
DC-DC
Converter
Output
Cable
shown
in
Figure
4-15
from
connector
\J504"
on
the A1
Main
Board
Assembly
.
3.
Turn the
4263B
on.
4.
Using
the
multimeter
,
check
if
the
DC-DC
Converter
open
output
voltages
on
the output
connector
pins
shown
in
Figure
4-15
are
in
the
ranges
shown
in
T
able
4-2.
5.
If
passed,
replace
the
A1
Main
Board
Assembly
,
otherwise
replace
the
DC-DC
Converter
Assembly
.
Figure
4-15.
DC-DC
Converter
Output
Connector
T
able
4-2.
DC-DC
Converter Open
Output V
oltages
Pin
(+)
Pin
(
0
)
V
oltage Range
1
2
5.1
6
0.2
V
4 3 12
6
0.6 V
5 3
0
12
6
0.6 V
4-18 Troubleshooting