5 Operating in EI Mode
To Autotune the MS for EI Mode
118 7000/7010 Series TQ GC/MS Operating Manual
9 Click Autotune to start the autotune. The Status line displays the current step
in the autotune process, and the plot of the tuned parameter for that step is
shown in the top graph. If specified above, at the completion of the autotune,
a Tune Report is printed.
To stop the autotune before it completes the automatic parameter selection,
click Abort Autotune. The parameters from the last successful autotune are
used.
10 Review the Tune Report. If the results are acceptable and you did not select
the Save tune file when done check box, save the autotune by clicking the
Files and Report tab, then click Save.
Refer to the manuals or online help provided with your Agilent MassHunter
GC/MS Acquisition software for additional information about tuning.
Figure 38. Setting up for autotune on a system configured with CI