8CI Maintenance
To Remove the CI Source Filament
238 7000/7010 Series TQ GC/MS Operating Manual
To Remove the CI Source Filament
Materials needed
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Tweezers (8710-2460)
Procedure
1 Vent the MS. (See “To Vent the MS” on page 124.)
2 Open the front analyzer chamber. (See To Open the Front Analyzer
Chamber” on page 156.)
3 Remove the CI source. (See “To Remove the 7000 Series CI Source” on
page 220.)
4 Remove the screw holding the filament to the CI source body. (See Figure 80
on page 239.)
5 Slide the filament off the CI source assembly. (See Figure 80 on page 239.)
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
The analyzer, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.