7 General Maintenance
To Clean the EI HES
7000/7010 Series TQ GC/MS Operating Manual 171
To Clean the EI HES
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder, 100 g (393706201)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Solvents
Methanol (reagent-grade)
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Ultrasonic bath
Procedure
1 Disassemble the ion source. (See “To Disassemble the EI HES” on page 165
or “To Disassemble the EI XTR Source” on page 168.)
2 Collect the following parts to be cleaned if you are cleaning an EI HES: (See
Figure 54 on page 172.)
Repeller
Source body
Extractor lens (5)
Post extractor lens 1 (4)
Post extractor lens 2 (3)
Ion focus lens (2)
Extended entrance lens (1)
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Series and is the answer not in the manual?

Agilent Technologies 7000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Series
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals