4 Installing 7890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using a Standard Column Nut
86 7000/7010 Series TQ GC/MS Operating Manual
To Install a Capillary Column in the GC/MS
Interface Using a Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MS
interface: The standard column nut explained here, and the self-tightening
column nut explained in the previous section.
Materials needed
Interface tip seal (G3870-20542) (See Figure 20 and Figure 21 on page 83.)
Tip seal spring (G7005-20024)
Knurled tip seal retainer (G3870-20547)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Interface column nut (05988-20066)
Ferrules
0.3 mm id, for 0.10 mm id columns (5062-3507)
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
0.5 mm id, for 0.32 mm id columns (5062-3506)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Series and is the answer not in the manual?

Agilent Technologies 7000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Series
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals