7 General Maintenance
To Clean the EI XTR Source
174 7000/7010 Series TQ GC/MS Operating Manual
To Clean the EI XTR Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder, 100 g (393706201)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the EI XTR source. (See “To Disassemble the EI XTR
Source” on page 168.)
2 Collect the following parts to be cleaned for an EI XTR source: (See Figure 55
on page 175.)
Repeller
Repeller block insert
Entrance lens
Ion focus lens
Extractor lens
Source body
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.