8CI Maintenance
To Disassemble the CI Source
7000/7010 Series TQ GC/MS Operating Manual 229
To Disassemble the CI Source
Materials needed
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Nut driver, 5.5 mm (8710-1220)
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view and the CI source parts list while using this
procedure. (See Figure 77 and Table 19 on page 230.)
1 Remove the CI source. (See “To Remove the 7000 Series CI Source” on
page 220.)
2 Remove the filaments. (See “To Remove the CI Source Filament” on
page 238.)
3 Separate the source heater assembly from the source body. The source
heater assembly includes the source heater, repeller, and related parts. (See
Figure 77 and Table 19 on page 230.)
4 Disassemble the repeller assembly by removing the ceramic insulator from
the repeller. (See Figure 77 on page 230.)
5 Remove the setscrew securing the lenses to the source body.
6 Pull the lenses out of the source body and separate the lens insulator, ion
focus lens, drawout cylinder, drawout lens, and entrance lens. (See Figure 77
on page 230.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Series and is the answer not in the manual?

Agilent Technologies 7000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Series
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals