6 Operating in CI Mode
Setting Up Your MS to Operate in CI Mode
130 7000/7010 Series TQ GC/MS Operating Manual
Setting Up Your MS to Operate in CI Mode
Setting up your MS for operation in CI mode requires special care to avoid
contamination and air leaks.
Always use the highest purity methane (and other reagent gases, if
applicable). Methane must be at least 99.9995% pure.
Always verify the MS is performing well in EI mode before switching to CI.
Ensure the CI source and GC/MS interface tip seal are installed.
Ensure the reagent gas plumbing has no air leaks. This is determined in PCI
mode, checking for m/z 32 after the methane pretune.
Ensure the reagent gas inlet line(s) are equipped with gas purifiers (not
applicable for ammonia).