8CI Maintenance
To Remove the CI Source Radiator from the 7010 Series TQ
7000/7010 Series TQ GC/MS Operating Manual 227
To Remove the CI Source Radiator from the
7010 Series
TQ
Materials needed
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Tweezers (8710-2460)
• T10 Torx screwdriver (5182-3466)
Procedure
This procedure is only applicable if you are using the 7010 side plate.
1 Remove the CI source. (See “To Remove the 7010 Series CI Source” on
page 221.)
The analyzers, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
When disconnecting leads, pull on the connectors, not on the wires.