EasyManuals Logo

Agilent Technologies 7000 Series User Manual

Agilent Technologies 7000 Series
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
6 Operating in CI Mode
To Operate the Reagent Gas Flow Control Module
140 7000/7010 Series TQ GC/MS Operating Manual
2 Use the parameters in the CI Reagent Gas Control area to control reagent gas
flow.
CI Gas A - Selects Methane as the reagent gas.
CI Gas B - Selects the gas connected to the B inlet on the gas flow controller
as the reagent gas.
CI Gas Flow - Enter percent of maximum volumetric flow for the selected
reagent gas. The actual % transmitted by the flow controller is shown next
to this entry. A good value for PCI is 20%, and 40% is a good value for NCI.
Set - Opens the selected reagent gas supply valve and controls the reagent
gas flow to the entered setpoint.
Off - Turns off the reagent gas flow.
Purge - Opens the selected reagent gas valve for 6 minutes to clear the
system of unwanted compounds.
Pumpout - Closes both reagent gas valves for 4 minutes, and evacuates the
system of reagent gases. At the end of the pumpout time, the selected
reagent gas valve opens.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Series and is the answer not in the manual?

Agilent Technologies 7000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Series
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals