Installation
8-7
Environmental
Conditions
The ACC mounting location must have adequate ventilation and air
circulation, and be in a relatively dust/lint-free, static-free environment.
The operating environmental characteristics of the ACC mounting location
must conform to the environmental specifications listed.
Temperature Range (ambient): Operating 0
o
C to 55
o
C (32
o
F to 131
o
F)
Humidity: Up to 95% RH
Altitude: Operating to 4600 meters (15,000 ft.)
Important Instructions
Allow a minimum of 5.08 cm (2 in.) of clearance on top and bottom.
Allow a minimum of 1 m (3 ft.) of free space in for service access.
Do not mount in an unventilated area such as a closet or console.
Mount within 1.5 m (5 ft.) of an A.C. power source.
When disconnecting from power, pull the cord cap.
Power Requirements:
100 - 240V at 50 - 60 Hz
Power Consumption:
80 VA.
Heat Dissipation:
51.6 kg/cal/hr (204.8 BTU/hr.).
Line Conditioning Requirements: None
•Weight:
14.1 lb. (6.4 kg).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 78581B and is the answer not in the manual?

Agilent Technologies 78581B Specifications

General IconGeneral
BrandAgilent Technologies
Model78581B
CategoryController
LanguageEnglish

Related product manuals