SDN/ACC Specifications
10-3
ACC Specifications
Table 10-2: ACC Specifications
–Class I
– Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide.
– IPX0 enclosure not protected against ingress of liquids.
– Mode of operation: Continuous
Table 10-3: Physical Characteristics
Electrical Characteristics Value
Power Requirements 100 to 240 VAC -10% to +6% 50-60 Hz
Power Consumption 80 VA
Heat Dissipation 51.6 kg/cal/hr
(204.8 BTU/hr)
Line Conditioning None
Leakage Current Less than 300 microamps
Physical Characteristics Value
Dimensions 1.735" (4.41 cm) x 17.665" (44.87 cm) x
17.25" (43.8 cm) (without wings)
Dimensions of optional
wall mount assembly
2.5" (6 cm) x 20.0" (51 cm) x 17.4" (44 cm)
Weight 14.1 pounds (6.4 kg)
Weight with wall mount
assembly
24.2 pounds (11.0 kg)
Service Access Ability to pull ACC forward in the rack-
mounted option leaving power cord and
secondary ground cable long. LEDs are
visible with cover on, but replacement
requires removal.