EasyManua.ls Logo

Agilent Technologies 78581B - ACC Specifications

Agilent Technologies 78581B
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SDN/ACC Specifications
10-3
ACC Specifications
Table 10-2: ACC Specifications
–Class I
Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide.
IPX0 enclosure not protected against ingress of liquids.
Mode of operation: Continuous
Table 10-3: Physical Characteristics
Electrical Characteristics Value
Power Requirements 100 to 240 VAC -10% to +6% 50-60 Hz
Power Consumption 80 VA
Heat Dissipation 51.6 kg/cal/hr
(204.8 BTU/hr)
Line Conditioning None
Leakage Current Less than 300 microamps
Physical Characteristics Value
Dimensions 1.735" (4.41 cm) x 17.665" (44.87 cm) x
17.25" (43.8 cm) (without wings)
Dimensions of optional
wall mount assembly
2.5" (6 cm) x 20.0" (51 cm) x 17.4" (44 cm)
Weight 14.1 pounds (6.4 kg)
Weight with wall mount
assembly
24.2 pounds (11.0 kg)
Service Access Ability to pull ACC forward in the rack-
mounted option leaving power cord and
secondary ground cable long. LEDs are
visible with cover on, but replacement
requires removal.

Table of Contents

Related product manuals